C-Link Protocol Commands Alarms alarm internal temp min alarm internal temp max These commands report the internal temperature alarm minimum and maximum value current settings. The example that follows reports that the internal temperature alarm minimum value is 15.0 °C. Send: Receive: internal temp alarm min internal temp alarm min 15.0 deg C set internal temp alarm min value set internal temp alarm max value These commands set the internal temperature alarm minimum and maximum values to value, where value is a floating-point number representing internal temperature alarm limits in degrees C. The example that follows sets the internal temperature alarm maximum value to 45.0 °C. Send: Receive: set internal temp alarm max 45 set internal temp alarm max 45 ok alarm pressure min alarm pressure max These commands report the pressure alarm minimum and maximum value current settings. The example that follows reports that the pressure alarm minimum value is 205 mmHg. Send: Receive: pressure alarm min pressure alarm min 205 mmHg set alarm pressure min value set alarm pressure max value These commands set the pressure alarm minimum and maximum values to value, where value is a floating-point number representing pressure alarm limits in millimeters of mercury. The example that follows sets the pressure alarm maximum value to 215 mmHg. Send: Receive: Thermo Fisher Scientific set alarm pressure max 215 set alarm pressure max 215 ok Model 43i Instruction Manual B-13