Download Model 17i - User Manual
Transcript
Servicing Photomultiplier Tube Replacement 9. Re-install the measurement bench. Refer to “Removing the Measurement Bench and Lowering the Partition Panel” in this chapter. Photomultiplier Tube Replacement Use the following procedure to replace the PMT tube. Equipment Required: Photomultiplier tube and PMT base Nut driver, 5/16-inch Flat blade screwdriver Philips screwdriver, small Equipment Damage Some internal components can be damaged by small amounts of static electricity. A properly grounded antistatic wrist strap must be worn while handling any internal component. ▲ 1. Turn instrument OFF, unplug the power cord, and remove the cover. 2. Disconnect the high voltage cable from the PMT power supply and unplug the BNC cable from the Input Board. 3. Remove six external screws holding PMT cover plate and the four screws holding the PMT shroud to the panel and remove the PMT cover plate (Figure 7–9). If the cooler fan is attached, unplug the fan power cord if necessary. PMT Cover Plate Screws (6) PMT Shroud Screws (4) PMT Base Screws (3) PMT Base Assembly Figure 7–9. Replacing the PMT Thermo Fisher Scientific Model 17i Instruction Manual 7-19