Download Model 17i - User Manual
Transcript
Geysitech (Bayern-Hessen) Protocol Geysitech Commands Operating and Error Status See Table D–1 for operating status and Table D–2 for error status for the Model 17i. Table D–1. Operating Status for Model 17i D7 D6 D5 D4 D3 D2 D1 D0 Bit 8 7 6 5 4 3 2 1 Hex-value 80 40 20 10 08 04 02 01 MSB LSB Operating status: Service Mode (On) 0 0 0 0 0 0 0 1 Maintenance (Local) 0 0 0 0 0 0 1 0 Zero gas (On) 0 0 0 0 0 1 0 0 Span gas (On) 0 0 0 0 1 0 0 0 Gas Unit Indication (ppm OR ppb) 0 0 0 1 0 0 0 0 Ozonator (Off) 0 0 1 0 0 0 0 0 PMT (Off) 0 1 0 0 0 0 0 0 Not used 1 0 0 0 0 0 0 0 Table D–2. Error Status for Model 17i D7 D6 D5 D4 D3 D2 D1 D0 Bit 8 7 6 5 4 3 2 1 Hex-value 80 40 20 10 08 04 02 01 MSB LSB Error status: Thermo Fisher Scientific Not Used 0 0 0 0 0 0 0 1 Not Used 0 0 0 0 0 0 1 0 Not Used 0 0 0 0 0 1 0 0 Any Temperature Alarm 0 0 0 0 1 0 0 0 Pressure Alarm 0 0 0 1 0 0 0 0 Sample Flow Alarm 0 0 1 0 0 0 0 0 Ozonator Flow Alarm 0 1 0 0 0 0 0 0 Not used 1 0 0 0 0 0 0 0 Model 17i Instruction Manual D-7