Download User`s Manual

Transcript
83
<10. Explosion Protected Type Instrument>
10.2FM
(1) Technical Data
(AXF Integral Flowmeter), (AXF Remote Flowtube)
Applicable Standard:
FM3600, FM3610, FM3615,
FM3810, ANSI/NEMA 250
(AXF Integral Flowmeter)
Explosion proof for Class I, Division 1, Groups A, B, C &
D.
Dust-ignition proof for Class II/III, Division1, Groups E, F
& G.
Intrinsically safe (electrodes) for Class I, Division 1,
Groups A, B, C & D.
“SEAL ALL CONDUITS WITHIN 18 INCHES”
“WHEN INSTALLED IN DIV. 2, SEALS NOT REQUIRED”
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: NEMA 4X
Temperature Code: T6
(AXFA14 Remote Converter)
Applicable Standard:
FM3600, FM3615, FM3810, ANSI/NEMA 250
Explosion proof for Class I, Division 1, Groups A, B, C &
D.
Dust-ignition proof for Class II/III, Division 1, Groups E, F
& G.
“SEAL ALL CONDUITS WITHIN 18 INCHES”
“WHEN INSTALLED IN DIV. 2, SEALS NOT REQUIRED”
Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: NEMA 4X
Temperature Code: T6
Ambient Temp.: –40°C to +60°C (–40°F to +140°F)
(2) Electrical Connection
The type of electrical connection is stamped near the
electrical connection port according to the following
codes.
(AXF Integral Flowmeter)
Screw Size
Marking
ISO M20x1.5 female
M
ANSI 1/2NPT female N
Note: Temperature Code T5 to T3 included in the scope of
application and its approval
Refer to following table;
Temperature
Code
T6
T5
T4
T3
Maximum Process
Temperature
+70°C (+158F)
+85°C (+185F)
+120°C (+248F)
+130°C (+266F)
Minimum Process
Temperature
–40°C (–40°F)
–40°C (–40°F)
–40°C (–40°F)
–40°C (–40°F)
Ambient Temp.: –40°C to +60°C (–40°F to +140°F)
(AXF Remote Flowtube)
Explosion proof for Class I, Division 1, Groups A, B, C &
D.
Dust-ignition proof for Class II/III, Division1, Groups E, F
& G.
Intrinsically safe (electrodes) for Class I, Division 1,
Groups A, B, C & D.
“SEAL ALL CONDUITS WITHIN 18 INCHES”
“WHEN INSTALLED IN DIV. 2, SEALS NOT REQUIRED”
Electrode Circuit Um: 250 Vac/dc
Excitation Circuit: 170V max
Enclosure: NEMA 4X
Temperature Code: T6
(AXF Remote Flowtube)
Screw Size
Marking
ISO M20x1.5 female
M
ANSI 1/2NPT female N
F1009.ai
(AXFA14 Remote Converter)
Screw Size
Marking
ISO M20x1.5 female
M
ANSI 1/2NPT female
N
10
Explosion Protected Type Instrument
Note: Temperature Code T5 to T3 included in the scope of
application and its approval
Refer to following table;
Temperature
Code
T6
T5
T4
T3
Maximum Process
Temperature
+70°C (+158°F)
+85°C (+185°F)
+120°C (+248°F)
+150°C (+302°F)
Minimum Process
Temperature
–40°C (–40°F)
–40°C (–40°F)
–40°C (–40°F)
–40°C (–40°F)
F1010.ai
Ambient Temp.: –40°C to +60°C (–40°F to +140°F)
IM 01E20A01-01EN