Download 1 Contents Chapter 1 Introduction
Transcript
58 CHAPTER 7 Maintenance Code Description Probable Cause Solution 16 An appropriate delivery system is not in place. Verify that the delivery system uses a Convergent compatible connector. Fiber is faulty or heavily damaged. Replace fiber. Remote connector missing. Verify that the remote connector, located on the back panel, is seated securely. An external device caused the remote interlock circuit to disable the laser system. Re-engage the remote interlock with the external source 17 18 27 Delivery system not connected Remote interlock open Foot switch not connected Blast shield error Verify that the connector is screwed down finger tight. An appropriate foot switch Verify that the foot switch is not in place. connector, located on the back panel, is seated securely. The foot switch connector is damaged. Call for service. Blast shield motor or optical sensor error. Call for service. Blast Shield installed backwards. Call for service. 28 Laser output below minimun Laser resonator out Call for service. of alignment or laser resonator mirror damaged. 29 Laser output 100% above set point System requires re-calibration. See Power Calibration in this chapter. Error spike injected into high power electronics. Turn OFF laser system and restart after 30 seconds. Odyssey 30 B User’s Manual