Analysis mode, Acquisition mode or Instrument off).
Figure 6.6 Switching the instrument into acquisition mode.
2) After the instrument has been primed and all system checks have been
performed successfully, the instrument status and LEDs will display green (see
section 6.3.5). If the system status is red, act in accordance with the error
message. The system status will prompt you if the instrument needs to be
calibrated and how many days since the last calibration.
3) If the LEDs are red or if you suspect one of the solutions is too low, you can
replace the appropriate solution.
4) To replace a solution place the closed bottle into the orange solution basket
prior to exchanging it. Remove the cap, and attach the bottle closure and
sensor to the new bottle.
5) To empty the waste container, remove the closure while the container is still in
the holder, close it with a cap. Then remove the closed container. Replace the
bottle with an empty waste bottle and attach the bottle closure and sensor to
the new waste bottle.
Note: Handle the full liquid waste bottle with extreme caution and dispose of as
recommended by your local authority.
Note: It is recommended to add 100 mL of a MACS Bleach solution to the bottom of
the waste container.
Note: Check that all connections are securely fastened and that no tubing is tangled.
6.3.5 Check the instrument status
The instrument status can be monitored using the status bar and illuminated bottles.