Download 959 Hot Cathode Controller Operation and

Transcript
13.6.3 Protect Pressure - PRO
Description:
Assigns the protection pressure for the Hot Cathode
sensor. This is the value at which the Hot Cathode
sensor’s filament will automatically have power
removed from it so that it can be protected from
exposure to high pressures. At pressures above
this value, power is removed from the Hot Cathode
filament and can not be reapplied by the user; at
pressures below this value, power may applied to
the Hot Cathode filament. The query response
returns the current protect pressure. A default
protect pressure exists at 1x10-2 Torr. The PRO
command overrides the default protect pressure
with a user selectable value that may be more
appropriate for the user’s application. The protect
pressure is not user assignable when the Controller
is in the Combined or Control Set Point mode of
operation.
Keyword:
PRO
Type:
Command and Query
Parameter:
numerical value expressed in scientific notation
Limits:
Min = 1x10-6 Torr
Max = 1x10-2 Torr.
Reset Value:
1.0E-2 Torr
Query Response:
numerical value expressed in scientific notation
Related Keywords:
SPx, CSP
Examples
Query:
Query Response:
Command:
Query:
Query Response:
PRO?;
ACK1.0E-2;
PRO!2.0E-3;
PRO?;
ACK2.0E-3;
959 Hot Cathode Controller