Download A35 - Ortec
Transcript
ScintiVision®-32 v2 (A35-B32) LOCK “Pwd” [,”Name”] This locks the current Detector using “Pwd” as the password. If the optional “Name” parameter is missing, the Locked name defaults to “Job”. This password is retained in the .JOB file and used with any .JOB commands so that the user does not need to re-enter the password. LOOP <repetitions> ... END_LOOP This pair executes multiple times all the commands between LOOP and END_LOOP. The number of execution times is specified by <repetitions>. Each command must be given on a separate line. A value of 0 executes once. A LOOP with no END_LOOP statement executes once. The loop count value can be included in any text by typing three question marks in the text where the loop count is to be inserted. The loop count replaces “???” with three characters using leading zeros if necessary. The loop variables, $(Loop) and $(LOOP1), can be included in any text. The loop count will be inserted with leading zeros suppressed. The following is an example: SET_DETECTOR 1 SET_PRESET_LIVE 20 LOOP 3 SET_DETECTOR 1 CLEAR START WAIT 2 WAIT FILL_BUFFER SET_DETECTOR 0 SAVE “TEST???.CHN” END_LOOP The above commands run three 20-second acquisitions and store the data on a disk in TEST001.CHN, TEST002.CHN and TEST003.CHN. If the SAVE command is replaced with SAVE “TEST$(Loop).CHN,” then the following files will be saved: TEST0.CHN, TEST1.CHN, and TEST2.CHN. 228