APPENDIX C. JOB FILE ERROR MESSAGES
No more unknown peaks!
The unknown peak finding sidebar function in Analysis Results display mode cannot find
any more unknown peaks in the direction indicated.
No peaks found!
The peak search function could not find any valid peaks in the spectrum.
No ROI found to report!
The REPORT function could not find any ROIs in range.
No ROI There To Clear.
The ROI Clear function (DEL key) requires at least one channel at the marker with the ROI
bit set.
No segment selected.
The Detector function could not be performed because no Segment was selected.
No Start Channel Was Defined.
The ROI “End” function was attempted without properly performing the “Begin” function.
Not Allowed During Acquisition!
An execution error (micro 135) arising from the Detector, indicating that the Detector
command is not allowed while acquisition is in progress.
Not Allowed During Current Mode!
An execution error arising from the Detector (micro 136), indicating that the Detector
command attempted is not allowed in the current mode of operation.
Not enough data points for fit.
Efficiency calibration fit for specified type requires a minimum number of table entries for
that type; e.g., for a polynomial fit, six or more points are required.
Not enough memory -- STRIP aborted.
The STRIP function temporarily allocates enough memory to read the file, but the allocation
failed in this case, probably due to insufficient available memory in the system. The STRIP
function is discontinued.
Not enough memory for COMPARE.
The COMPARE mode could not be executed due to insufficient memory for the second
spectrum.