Transcript
LCD Display: Alert: Cause: Resolution: ALERT NOT in PCA. NO BOLUS ALLOWED Bolus requested when therapy is not PCA Remote Bolus button or BOLUS key is pressed when not in PCA therapy. Remove the remote bolus cord from the pump if connected. Refrain from pressing the BOLUS key. LCD Display: Alert: Cause: ALERT NUMBER OF DOSES INVALID Number of doses entered exceeds maximum for therapy The number of doses per bag in Intermittent therapy exceeds the allowable maximum. Recheck data entry. The number of doses requested must be less than or equal to the maximum for the therapy. Resolution: LCD Display: Alert: Cause: Resolution: LCD Display: Alert: Cause: Resolution: LCD Display: Alert: Cause: Resolution: LCD Display: Alert: Cause: Resolution: ALERT PERFORM ROUTINE MAINTENANCE Maintenance Date (displayed only if date has arrived when pump is turned on) Current Date is beyond the Maintenance Date. Return pump for routine maintenance. ALERT RAMP INVALID FOR AMT/RATE Ramp time entered is invalid for amt/rate already programmed Ramp value entered for TPN is unacceptable given the desired rate and amount programmed. Recheck all data entered for prescription. If programming a new TPN therapy (”NEW PROGRAM” selected) and the ramp time is a requirement, enter the ramp time(s) and total time first followed by the next governing parameter (rate or volume to be infused). ALERT RATE EXCEEDS DELTA RATE/DELTA TIME Rate entered exceeds the delta time limits The rate entered falls outside of the allowable range for the delta rate / delta time limits programmed. Recheck data entry or wait more time elapses then attempt to change the rate. Rate selected cannot exceed the range placed on titration. ALERT RUNNING, PAUSE PUMP BEFORE OFF Pause Infusion before turning off The OFF key was pressed while the pump was infusing or a running screen displayed. Press “PAUSE” key before pressing “OFF” key to stop an infusion. 128