Download WINWCP User Guide - Spider

Transcript
Signal Averaging > Introduction
Many electrophysiological signals have poor signal-noise ratios, making it difficult to obtain accurate measurements from
individual records. However, if a signal can be made to occur repeatedly, digital signal averaging techniques can recover
the signal waveform from the background noise.
The signal average of a series of records is generated by computing the average of each corresponding sample within the
records. For a set of N records, consisting of samples, yi, (i=1,n) the average record consists of n samples, and is given
by.
N
Avg = N1 ∑ y
i
r =1
i
The location of signals within the record sometimes varies from record to record, due to imperfections in the detection of
spontaneous signals or fluctuations in stimulus latency. In such circumstances, averaging corresponding sample points
within the record would result in a distorted signal average. This problem can be avoided by aligning the signals by the
mid-points of their rising phases before averaging.
131