Download Model 48i High Level
Transcript
Operation Service Menu CALIBRATE FLOW ZERO: CURRENTLY: 1.139 LPM SET TO: 0.000 LPM ? DISCONNECT PUMP AND SAVE CURRENT FLOW RANGE Calibrate Flow Span AVG DIAGS ALARM The Calibrate Flow Span screen allows the user to view and set the flow sensor calibrate span point. Note An independent flow sensor is required to read the flow, then the operator enters the flow value on this screen to perform the calibration. ▲ ● In the Main Menu, choose Service > Flow Calibration > Span. CALIBRATE FLOW SPAN: CURRENTLY: 1.139 LPM SET TO: 1.141 LPM ? §¨ MOVE CURSOR ©ª CHANGE VALUE SAVE RANGE Restore Default Flow Calibration AVG The Restore Default Flow Calibration screen allows the user to reset the flow calibration configuration values to factory defaults. ● In the Main Menu, choose Service > Select Flow Calibration A or B > Set Defaults. RESTORE DEFAULT CAL: RESTORE RANGE 3-72 Model 48i High Level Instruction Manual DIAGS ALARM AVG DIAGS ALARM RESTORE DEFAULT CAL: RESTORE ARE YOU SURE YOU WANT TO?x PRESS ¨ TO CONFIRM RESTORE RANGE AVG DIAGS ALARM Thermo Electron Corporation