Operation
Service Menu
CALIBRATE FLOW ZERO:
CURRENTLY:
1.139 LPM
SET TO:
0.000 LPM ?
DISCONNECT PUMP AND
ƒ SAVE CURRENT FLOW
RANGE
Calibrate Flow Span
AVG
DIAGS ALARM
The Calibrate Flow Span screen allows the user to view and set the flow
sensor calibrate span point.
Note An independent flow sensor is required to read the flow, then the
operator enters the flow value on this screen to perform the calibration. ▲
●
In the Main Menu, choose Service > Flow Calibration > Span.
CALIBRATE FLOW SPAN:
CURRENTLY:
1.139 LPM
SET TO:
1.141 LPM ?
§¨ MOVE CURSOR
©ª CHANGE VALUE
ƒ SAVE
RANGE
Restore Default Flow
Calibration
AVG
The Restore Default Flow Calibration screen allows the user to reset the
flow calibration configuration values to factory defaults.
●
In the Main Menu, choose Service > Select Flow Calibration A or B >
Set Defaults.
RESTORE DEFAULT CAL:
ƒ RESTORE
RANGE
3-72
Model 48i High Level Instruction Manual
DIAGS ALARM
AVG
DIAGS ALARM
RESTORE DEFAULT CAL:
ƒ RESTORE
ARE YOU SURE YOU WANT TO?x
PRESS ¨ TO CONFIRM RESTORE
RANGE
AVG
DIAGS ALARM