Download Manual - Equipco
Transcript
Chapter 5 Troublesh~oting appears then replace the W lamp. See Section 4.4. Fault: Calibration error occurred error occurred Cause: Calibration gas is not being supplied correctly. Action: Ensure the calibration gas cylinder valve is open all the way. Action: Ensure there is calibration gas in the cylinder. Open the cylinder and check the contents gauge. Action: If pressurized calibration gas is beingused, ensure the delivery pressure is maintained a t 5 psig (35 kPa). Ensure the delivery pressure does not exceed 5 psig (35 kPa). If you are using the Photovac calibration gas regulator, i t has been preset to deliver 250 mumin. Use a flowmeter to check the flow rate. Action: If a gas sampling bag is being used, ensure the valve on the bag is open before connecting it to Snapshot. Action: Before beginning a calibration with the STARTfSTOPkey, turn on the calibration gas and allow the adapter tee and adapter tubing to be flushed with calibration gas. Action: There is a n undetermined problem. Contact the EQUIPCO Service Department a t (888)234-5678. Cause: Calibration gas does not contain all of the compounds for the application. Action: In order for Snapshot to calibrate itself correctly, all the compounds from the application module must be contained in the calibration gas. 15 ppa www.equipcoservices.com Snapshot User's Manual 51