Download Manual - Equipco
Transcript
Chapter 1 Introduction 9. After the lamp has been tuned, SnapShot will purge the column and heat the isothermal column oven. The instrument status will be "Chck" while the oven heats. The status will then change to T u r f while the column is purged for the duration of one analysis. When the instrument status changes to Tedy" Snapshot is ready for operation. A new cylinder will provide approximately 8 hours of operation. If the instrument is not in use you can stop the flow of carrier gas by turning the carrier gas shut-off valve clockwise. Snapshot will automatically turn the lamp off when the carrier gas supply is shut off. The carrier gas shut-off valve does not control the actual flow rate. The flow rates are set automatically by the application module. Flow rates will differ from one application module to another. Note: Do not remove the carrier gas cylinder from Snapshot until the cylinder is empty. You will know the cylinder is empty when a "Chck" status is displayed. Press the TUTOR key. The message will be "Carrier pressure is now low". www.equipcoservices.com Snapshot User's Manual 9