Download Test Tool User Manual
Transcript
Get Control Accelerometer Test Tool - Version 5.0 If the tool measures earth gravity to +/-0.1g (+/-10%) using the Calibration Display feature, it is working properly. To remove the system DC offsets to get a more accurate verification, use a level to improve the accuracy of each axis measurement. Record each positive and negative axis reading. Then subtract the negative axis measurement from the positive axis measurement and divide this result by 2. This number should be close to 1.00g. This method removes the tool’s DC offsets from the measurements. However, it does not account for geographical earth gravity variances or inaccuracies in orienting the tool perpendicular to earth gravity. 12