10.LIST OF PARAMETERS
No.
Item
01 Model
02 Tag no.
03 Self-check result
Display
MODEL
TAG NO
SELF CHK
A
A01
A05
A06
A07
A08
A17
A18
A20
A60
Display 1
Input value
Output value 1
Output value 2
Alarm-1 status
Alarm-2 status
Status
Rev. no.
Menu rev.
Self-check result
DISPLAY1
INPUT1
OUTPUT1
OUTPUT2
ALM1 STATUS
ALM2 STATUS
STATUS
REV NO
MENU REV
SELF CHK
D
D01
D02
D03
D04
D11
D12
D38
D39
D60
Setting (I/O)
Tag no. 1
Tag no. 2
Comment 1
Comment 2
Linearize
Low cut
Direction of output-1 action
Direction of output-2 action
Self-check result
SET(I/O)
TAG NO.1
TAG NO.2
COMMENT1
COMMENT2
LINEARIZE
LOW CUT
OUT1 DR
OUT2 DR
SELF CHK
There are items not displayed depending on what
output is specified.
P
P02
P03
P12
P13
P14
P15
P60
Adjusting items
Adjustment
ADJUST1
Zero adjustment of input 1
ZERO ADJ1
Span adjustment of input 1 SPAN ADJ1
0% adjustment of output 1
OUT1 0%
100% adjustment of output 1 OUT1 100%
0% adjustment of output 2
OUT2 0%
100% adjustment of output 2 OUT2 100%
Self-check result
SELF CHK
Remarks
No.
Item
Display
Display items
B Display 2
B01 Input value 1
B05 Output value 1
B06 Output value 2
B60 Self-check result
*1
*1
DISPLAY2
INPUT1
OUTPUT1
OUTPUT2
SELF CHK
The Status is displayed for service personnel to see history
records.
Setting items
E Setting (alarm output)
E01 Alarm-1 setting
E02 Alarm-2 setting
E03 Alarm-1 setting
E04 Alarm-2 setting
E05 Direction of alarm-1 action
E06 Direction of alarm-2 action
E09 Alarm-1 hysteresis
E10 Alarm-2 hysteresis
E11 Alarm-1 ON delay setting
E12 Alarm-2 ON delay setting
E13 Alarm-1 OFF delay setting
E14 Alarm-2 OFF delay setting
E15 Direction of alarm-1 relay action
E16 Direction of alarm-2 relay action
E60 Self-check result
F Setting (communication)
second
F01 Communication protocol
F02 Address
F03 Baud rate
F04 Parity
F05 Data length
F06 Stop bit
F60 Self-check result
Test items
Q Test
Q01 ON/OFF of RJC
Q02 Forced output 1
Q03 Forced output 2
Q04 Forced output (alarm 1)
Q05 Forced output (alarm 2)
Q60 Self-check result
11.MAINTENANCE
The product starts running immediately when the power is turned on;
however, it needs 10 to 15 minutes of warm-up before it meets the
specified performance.
For cleaning the instrument, use a soft and dry cloth.
11.1 Calibration Apparatus
● A voltage and current generator (Yokogawa 7651 or the
equivalent)
● A voltmeter (Yokogawa 7562 or the equivalent)
● A precision resistor of 250 ⍀ ± 0.01%, 1 W (for current output)
All Rights Reserved. Copyright © 1999, Yokogawa M&C Corporation
SET(ALM)
SET POINT1
SET POINT2
SET POINT1
SET POINT2
ALM1 ACTION
ALM2 ACTION
HYSTERESIS1
HYSTERESIS2
ON DELAY1
ON DELAY2
OFF DELAY1
OFF DELAY2
RL1 ACTION
RL2 ACTION
SELF CHK
SET(COM)
PROTOCOL
ADDRESS
BAUD RATE
PARITY
DATA LEN
STOP BIT
SELF CHK
TEST
RJC
OUT1 TEST
OUT2 TEST
ALM1 TEST
ALM2 TEST
SELF CHK
If the output signal is out of the rated accuracy range, adjust the
output signal level using the Handy Terminal (JHT200).
Input signal
4 to 20 mA DC input
+
Voltage and
current
generator
3
-
3
4
6
2
1
5
4
Output-2 signal
Analog output
2
5
L+
N-
GND
10
+
R - Voltmeter
R: For current output using 250 Ω
precision resistor
Power Supply
11.2 Calibration Procedure
Connect the instruments as shown in Fig. 10.1. First adjust the output-1 signal and then the output-2 signal.
Produce input signals equivalent to 0, 25, 50, 75, and 100% of the
input span (or those equivalent to 0, 6.25, 25, 56.25, and 100% of
the input span if square root extraction is set) from the voltage and
current generator to the distributor. Then, check that the
distributor’s output signal shows voltages corresponding to 0, 25,
50, 75, and 100% of the input span within the rated accuracy range.
Remarks
9
8
7
Output-1 signal
11 10
Analog output
11
7
8
9
+
R - Voltmeter
R: For current output using 250 Ω
precision resistor
Fig. 11.1
IM 77J01A07-01E