Download A Practical Guide to Flow
Transcript
Technical Note: A Practical Guide to Flow Measurement potential error to a negligible 0.3%. Remember that blanks should also be diluted 10-fold. Turbidity causes a second effect which is generally unrecognized and is a potential error, although only in rare situations. All glass and gelatin optical filters are somewhat fluorescent. Scattered excitation light strikes the emission filter and causes it to fluoresce. If the blank and standard are prepared from the sample water, this additional fluorescence simply yields a higher blank than the same water without turbidity and there is no error. If the dye concentration is low enough so that the added contribution is significant, then a change in turbidity will cause an error in the opposite direction from that previously discussed. The effect is usually small and seldom a problem with dye concentrations above one ppb (actual ingredient). It should be borne in mind, however, that some of the worst scattering can occur in the absence of visible particles. In the volumetric flume study (32), the injection rate was set to yield a rhodamine B concentration of one ppb at the highest flow (about 200 cfs). There was violent turbulence and many visible air bubbles in the early sections of the flume. Samples and blanks were collected far enough down the flume that the visible bubbles were absent. For convenience, however, water for the standards was collected from the river just prior to entry. At 200 cfs, the dye technique consistently calculated a flow 8% lower than the volumetric determination. The error reduced approximately linearly with reduction in flow. The problem was massive levels of micro air bubbles that were causing a blank increase of 0.08 ppb. In other words, the apparent dye concentration was 0.08 ppb higher than the true dye concentration. concentration of 10 ppb had been used in the beginning, the bias at 200 cfs would have been 0.8%, about 0.4% at 100 cfs, and would have probably been ignored. Unfortunately, sufficient dye was not available to make this increase after the problem was diagnosed. 4. The final problem is sample temperature. It is simplest to have your calibration standard and your samples at the same temperature. Float your carefully sealed standard in the stream. In the laboratory, place all the samples together, away from any source of heat. For precise studies, it is best to incubate both standard and samples in a water bath. The sample compartment of the fluorometer will be a few degrees above ambient temperature, but there will be no effect on the sample in the few seconds required for measurement. However, don’t attempt to remeasure any sample until it has re-equilibrated, because the cuvette will have begun to warm up. If you wish to periodically recheck samples, or for some reason wish to have them in the fluorometer for an extended period, use the 10-030 Sample Holder, which is temperature-regulated. If you need to correct various samples to a standard temperature, Wilson contains a simple graph from which an accurate correction is easily obtained (17). On the other hand, if you have a pocket calculator handy, the equation is: F r = Fs e [n(Ts - Tr)] Where: Fr Upon discovery, there was sufficient time to invent a practical means of removing the bubbles. Thus, water taken at the sample point was used to prepare standards and all standards, samples and blanks were measured 30 minutes after the water collection. This ploy was probably not totally successful due to the different treatment given to the standards (pipetting, mixing, etc.). (The researchers didn’t think to chill the samples, an ultrasonic bath was not available, and the director of the hydraulic laboratory indicated that it would take from 24-48 hours for the bubbles to disappear). Note that if a dye Sample Temperature is the calculated fluorescent reading at the reference temperature, Tr. Fs is the observed fluorescence reading of the sample at the time of reading the sample temperature, Ts. e is the base of natural log. n is a temperature coefficient for whatever dye is used. For Rhodamine WT, rhodamine B, pontacyl brilliant pink B, and fluorescein, the values have been www.turnerdesigns.com 998-5000 Revision A 05/31/2001 Page 17 of 29