GATE LED: it turns on with a green color when the SPAD is biased above breakdown; on the contrary it is inactive when the SPAD is biased below breakdown. This is true independently of the type of the used detection head (free-­‐‑running or gated). •
Detection Head LED: it turns red when there is no communication with the Detection Head or when the maximum Gate Width is not satisfied (external trigger). Otherwise it is green. •
Control Unit LED: it turns red only when there is no communication with the Control Unit. Otherwise it is green. •
System Ready LED: during the normal functioning of the whole system it is green; it turns red only when an error inside the module is detected. Others indicators in this section are: •
CU SN : MPD serial number of the connected Control Unit. •
DH SN : MPD serial number of the connected Detection Head. •
SPAD Temp : the current temperature of the detector in Kelvin. It is active only when the TEC controller is active; in all the other cases it does not show anything. There is also a status bar at the bottom of the module window (shown for example in Figure 2.10): it reports the module loaded configuration or eventually the error messages. Figure 2.13. Module Status Overview indicators, during detector power on. Micro Photon Devices S.r.l. – Italy