Power and Energy Measurement
BA500 User Guide
3. Select the version of firmware to install in the USBI device
4. Press the “Start” button. The “Done” button becomes inactive
until the upgrade is complete. The progress bar displays the
status of the upgrade. The upgrade takes approximately 2
minutes per device upgraded
5. At the end of the upgrade, press the “Done” button
6. If the USBI device does not respond immediately after the
upgrade, press the “Refresh Hardware” button in the RP-USB
display.
7.7.3
Zeroing
In the USBI, all adjustments, including zeroing internal circuits, are
performed from the software. This ensures simple and accurate
realignment. For best performance, it is recommended to zero the
USBI every 2 months.
To zero the instrument:
1. If a head is connected, disconnect the head, and press “refresh
hardware” to enable the USBI to identify that no head is
connected.
2. Check that the USBI is not in an electrically noisy environment
and is undisturbed.
3. Allow the USBI to run for at least 30 seconds.
4. Press the “Zeroing” button. This opens the Zeroing dialog box
50