Download USER MANUAL
Transcript
SECTION 2 INSTALLATION OF THE WATERLASE MDX INSTALLATION INSTRUCTIONS Your local authorized Biolase representative will unpack the Waterlase MDX and your authorized Biolase service representative will install the unit. Please leave all crates and shipping containers unopened until your authorized trained service representative arrives. Complete installation, testing and demonstration requires approximately one full day. The Waterlase MDX must be installed with an authorized Biolase representative; please refer to Section 4, Operating Instructions, for setup instructions. Please contact your authorized local Biolase representative before transporting unit to a different location. Misalignment of optical components may occur during transportation if the unit is not properly packaged. FACILITY REQUIREMENTS ● Electrical Supply 100 VAC @ 15.0 Amps to 230 VAC @ 8.0 Amps, 50/60 Hz ● Compressed Air Supply 80 - 120 psi (5.5 - 8.2 bar) CAUTION: Moisture in the air supply line may damage the laser system. Please provide proper filtration to eliminate all moisture from air source. Connections for air supply must be available in each operatory. Attach air hose with 1/4” inside diameter male quick connectors on each end between air inlet connector (fig. 3.2) and operatory air source. CAUTION: Prior to connection, verify that outlet is for air, not water supply. Connection to water supply may cause damage to the Waterlase MDX system. If the unit was connected to the water supply, do NOT turn the system on. Contact your authorized local Biolase service representative. 9