Download MODEL 700E DYNAMIC DILUTION CALIBRATOR
Transcript
TELEDYNE INSTRUMENTS M700E Calibrator Operator’s Manual General Troubleshooting & Repair of the M700E Calibrator 11.3.3. VALVE DRIVER PCA STATUS LEDS The Signal I/O submenu also includes VARS that can be used to turn the various input gas valves on and off as part of a diagnostic investigation. WATCHDOG INDICATOR CAL GAS VALVE 1 PURGE VALVE CAL GAS VALVE 2 DILUENT VALVE CAL GAS VALVE 3 CAL GAS VALVE 4 Figure 11-4: Valve Driver PCA Status LEDS Used for Troubleshooting Table 11-6: Valve Driver Board Watchdog LED Failure Indications LED Function Fault Status Indicated Failure(s) Failed/Halted CPU D1 (Red) I2C bus Health (Watchdog Circuit) Continuously ON or Continuously OFF Faulty Mother Board, Keyboard or Relay PCA Faulty Connectors/Wiring between Mother Board, Keyboard or Relay PCA Failed/Faulty +5 VDC Power Supply (PS1) Table 11-7: Relay PCA Status LED Failure Indications LED FUNCTION ACTIVATED BY SIGNAL I/O PARAMETER D3 Cal Gas CYL1 CYL_VALVE_1 D4 Cal Gas CYL2 CYL_VALVE_2 D5 Cal Gas CYL3 CYL_VALVE_3 Failed Valve D6 Cal Gas CYL4 CYL_VALVE_4 Failed Valve Driver IC on Relay PCA D9 Purge Valve Status PURGE_VALVE Failed Valve Driver Board D10 Diluent Valve Status INPUT_VALVE 05622 REV B DCN 5169 DIAGNOSTIC TECHNIQUE Valve should audibly change states and LED should glow. If not: Faulty +12 VDC Supply (PS2) Faulty Connectors/Wiring 247 PRINTED DOCUMENTS ARE UNCONTROLLED