Download Laser SL335
Transcript
33 8.5. & Primary water replacement Note: Here is a brief description; consult the Manual of cooling unit PS1222CO for details. 1. Pulling forwards, withdraw the cooling unit from Cabinet and remove the reservoir cork. By either siphon or stirrup-pump type device, drain the water from primary cooling contour into a suitable vessel. Refill the reservoir with distilled water (3.5 liters). 2. Inspect if the filter is not clogged. Replace it if necessary. 3. It is essential that none air be left in the primary circuit after the reservoir refill. To remove the air, turn the key [K1] and power [S3] switches to ON (Fig.3). Press button MEASURE/SET [B3], disabling all the blockings in the cooling unit. This should cause all air in cooling contour to be pumped off into the reservoir. In the case this did not help to remove air from pump, switch off power switch [S3], turn, by several rounds, the filter sump to let the air out of the internal contour. Turn the filter sump tightly again. When all the air is likely to have been removed, release the [B3] and check primary water level. Switch off the cooling unit ([S3] to OFF). Stuff the reservoir cork back and slide the unit into its place in the Cabinet. 8.6. Washing of cooling system In case you find the cooling unit infected with organic substance (kind of algae), the whole cooling system needs to be washed over. You will need: • 0.5 liter of 30% hydrogen peroxide or 0.5−1 litre of cleaning liquid Extran MA-03 phosphate-free (of Merck make); • 3.5 liter of distilled water; • a spare particle filter; • a spare DI filter. 1. Turn the laser system off. 2. Drain the water from cooling unit. 3. Fill the reservoir with distilled water adding the hydrogen peroxide or Extran. 4. Switch the cooling unit on. In dependence of the cleaning agent being used, leave the unit working: for an hour in case of hydrogen peroxide and for some 3−5 min in case of Extran. 5. Switch the unit off and drain its reservoir. 6. Replace the DI and particle filters with new. 7. Refill the reservoir with distilled water. 8. Switch the cooling unit on and make sure it works normally & Note. The temperature of cooling water in the cooling unit is pre-defined at ∼29°C. If the environmental temperature of your working place exceeds mentioned 29°C, the laser heads may start covering with dew. Therefore, the cooling water temperature needs to be. The maximum admissible temperature of cooling water is 32°C.