Procedure:
Note
Note
1. Set up the equipment as shown in Figure 10-2.
If the Interface Box is not available, use the following cables and
adapters as a substitute. Figure 10-3 shows the test setup without
the interface box.
BNC(m) to BNC(m) Cable, 30 cm Agilent PN 8120-1838 2ea.
Tee, BNC(m)(f)(f) Adapter
Agilent PN 1250-0781
2. Set the multimeter to ACV.
3. Perform a SYSTEM RESET as described in \System Reset".
4. Set 4284A's controls as follows:
Test Frequency: 1.25 kHz
High Power Option: OFF
5. Display the MEAS DISPLAY page.
6. Set the Oscillator Level in accordance with Table 10-3, and
conrm that the Multimeter reading and the Level Monitor
reading are within the test limits given in the table.
7. Set the Test Frequency to 960 kHz and perform step 6.
Steps 8 through 11 should be performed only when the 4284A is
equipped with Option 001.
8. Set the controls of the 4284A as follows:
Test Frequency: 1.25 kHz
High Power Option: ON
9. Display the MEAS DISPLAY page.
10. Set the Oscillator Level in accordance with Table 10-4, and
conrm that the Multimeter reading and the Level Monitor
reading are with in the test limits given in Table 10-3.
11. Set the Test Frequency to 960 kHz and perform step 10.
Performance Tests
10-7