Transcript
Troubleshooting Troubleshooting Using External Test Figure 4-47 Dynamic Range tests dialog box example (E5071B Opt. 413/414) b. Select the S parameters for the test by clicking the check boxes. Then, click “Start Test“ button. The instruction dialog box opens as shown in Figure 4-48. Follow the instruction. NOTE Figure 4-48 Clicking “Select All” button checks all the check boxes. Clicking "Select None" button clears all the check boxes. Click "Exit" to return to the External Test dialog box SHORT (all ports) dialog box c. Connect type-N Short terminations to all the test ports. NOTE For more information on the Dynamic Range tests setup, see Table 4-8. Table 4-8 Dynamic Range tests setup Test parameter Connect Short terminations to Sweep freq span, IFBW, Nop and RF power settings (Automatic) S12 S13 (Opt. 313/314/413/414) S14 (Opt. 413/414) S21 S23 (Opt. 313/314/413/414) S31 (Opt. 313/314/413/414) S32 (Opt. 313/314/413/414) S34 (Opt. 413/414) S41 (Opt. 413/414) S43 (Opt. 413/414) All the test ports 3 MHz to 3 GHz (E5070B), 248 3 MHz to 8.5 GHz (E5071B), 3 kHz BW, 101, max. source power Chapter 4