Download Plexus C1000D SERIES

Transcript
Plexus C/1000D
9.0
Function Test Check Sheet
Model
Serial
Test Procedure Results
(see routine maintenance and function test section of manual for details)
Physical Inspection OK
Yes or No
Testing
All C/1000D
1.
2.
3.
High pressure reading observed in step F.
Low pressure reading observed in step F.
Flow test
_______ mmHg
_______ mmHg
Pass or Fail
Float / Power Fail options only.
4.
5.
Pressure on both gauges?
Power fail alarms function?
Pass or Fail
Pass or Fail
Electrical safety inspection results.
1.
Max current leakage reading ____ µ amps
2.
Ground resistance less than .5 ohms
Pass
Fail
Tested by _______________________________ Date ______________
If the unit passes all the steps described in the physical inspection, testing and electrical safety inspection
sections above it should be considered ready for use. Any unit that does not pass all the requirements above