GE MEDICAL SYSTEMS
DIRECTION 2395907, REVISION 2
8-13-1-5
LOGIQ™ 5 EXPERT BASIC SERVICE MANUAL
BEP2 Cable Connection with Console (cont’d)
6.) Connect LI, LO, MI, ETHERNET Connector to Sound/LAN Port on BEP2.
Ethernet Cable
MI (MIC : Red) cable
LO (Line Out : Green) Cable
LI (Line In : Blue) Cable
Figure 1-6 Connecting Cables
7.) Connect CON17 Connector to BEP2.
CON 17 cable
Figure 1-7 Connecting CON17 cables
8.) Check the CON 17 cable is arranged without short before connecting.
Chapter 8 - Replacement Procedures