Download Technical Publication
Transcript
GE MEDICAL SYSTEMS DIRECTION 2395907, REVISION 2 10-7-3 LOGIQ™ 5 EXPERT BASIC SERVICE MANUAL Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated. Any problems found should be reported to the hospital immediately and the receptacle should not be used. Figure 10-1 Typical Outlet Tester NOTE: 10-7-4 No outlet tester can detect the condition where the Neutral (grounded supply) conductor and the Grounding (protective earth) conductor are reversed. If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected. Grounding Continuity CAUTION Electric Shock Hazard. The patient must not be contacted to the equipment during this test Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case. The ground wire resistance should be less than 0.2 ohms. Reference the procedure in the IEC 601-1.1. Figure 10-2 Ground Continuity Test 10 - 12 Section 10-7 - Electrical Safety Tests