Download Manual

Transcript
Operating
Manual
Citizen Scale, Inc. USA
Call Toll Free Support : 1.800.997.1440
Visit : www.citizeninc.com
Weighing Indicator
CTB
CTB weighing indicator operation manual
REV C1, June 2012
CTB weighing indicator operation manual
CTB weighing indicator operation manual
REV C1, June 2012
10. CALIBRATION
The scale can be calibrated using the following procedure.
For a more
detailed method of calibrating the scale it will be necessary to enter the
secure Programming Menu. Refer to the CTB Service Manual for more
information.
The CTB scales calibrate using metric weights and the CTBa scales can
calibrate using either metric or pound weights, depending on the
weighing unit in use before calibration.
The display will show either
"kg" or "lb" to identify the weights expected.
Turn the power off. Turn the power back on, during the counting from 9 to
0 press the SET key.
The display will show "CAL " for a few seconds. While it is showing
"CAL " press the SMPL, PRINT and TARE keys in sequence to enter
the Calibration section. The display will show "unLoAd".
Remove any weight from the platform. Press the ZERO/ENTER key.
The display will show "LoAd". Place the calibration weight on the scale.
Press the ZERO/ENTER key.
The display will show the last calibration weight used. If this is correct
you can continue by pressing the ZERO/ENTER key.
If it is not correct
use the arrow keys to change the calibration weight value. When it is
correct press the ZERO/ENTER key.
If the calibration is acceptable the display will show "PASS" and then
return to normal. If an error message is shown try calibration again as a
disturbance may have prevented a successful calibration.
If the problem persist then contact Taiwan scale or your dealer.
After calibration the scale should be checked to verify the calibration and
linearity is correct. If necessary repeat calibration, especially be certain
the scale is stable before accepting any weight.
-10-
CTB weighing indicator operation manual
CONTENTS
1. Introduction………………………………………………………………….1
2. Key description…………………………………………………………….1
3. Basic operation…………………………………………………………….2
3.1 Zero the display…………………………………………………………..2
3.2 Tare……………………………………………………………………….2
3.3 Weighing a sample………………………………………………………3
4. Parts counting……………………………………………………………3
5. Check weighing…………………………………………………………..4
6. Accumulation………………………………………………………………5
7. Parameters……………………………………………………………..5
Function menu setting table…………………………………………6
8. Battery operation………………………………………………………….8
9. Error codes……………………………………………………………..9
10. Calibration………………………………………………………………..10
CTB weighing indicator operation manual
CTB weighing indicator operation manual
REV C1, June 2012
9. ERROR CODES
ERROR
CODES
-----
DESCRIPTI
ON
Over range
Err 1
Date Setting
Error
Err 2
Time Setting
Error
Err 4
Zero Setting
Error
Err 6
A/D out
range
Err 9
Unstable,
can’t return
to zero
of
RESOLUTION
Remove weight from the scale.
If the problem persist contact your dealer or
Taiwan scale for assistance.
Enter date using correct format and
reasonable values.
Format: yy:mm:dd
Enter time using correct format and
reasonable values.
Format: hh:mm:ss
The scale was outside the normal zero setting
range either when it was turned on or when
the ZERO key was pressed.
Remove weight from the scale and try again.
Use the TARE key to set the display to zero
value.
If the problem persist contact your dealer or
Taiwan scale for assistance.
The values from the A/D converter are outside
the normal range. Remove weight from the
scale if overloaded, make sure the pan is
attached.
Indicates the load cell or the electronics may
be faulty.
If the problem persist contact your dealer or
Taiwan scale for assistance.
When turn on the power, if internal counts is
not stable, display will have “Err 9”, please
check the platform and load cell.
If the problem persist contact your dealer or
Taiwan scale for assistance.
-9-
CTB weighing indicator operation manual
REV C1, June 2012
8. BATTERY OPERATION
The weighing indicator can be operated from the battery if desired.
The battery life is approximately 100 hours.
When the battery needs charging a symbol on the weight display will
turn on. The battery should be charged when the symbol is on. The
scale will still operate for about 10 hours after which it will
automatically switch off to protect the battery.
To charge the battery simply plug into the mains power. The scale does
not need to be turned on.
The battery should be charged for 12 hours for full capacity.
Just under the weight display is an LED to indicate the status of battery
charging. When the scale is plugged into the mains power the internal
battery will be charged. If the LED is green the battery has a full charge.
If it is yellow the battery is nearly full of charge and red indicates the
battery is being charged.
As the battery is used it may fail to hold a full charge. If the battery
life becomes unacceptable then contact your distributor
CTB weighing indicator operation manual
1. INTRODUCTION
The T2000A weighing indicator provides an accurate, fast and
versatile series of general purpose weighing indicator with counting and
check-weighing functions.
All the keypads are sealed, color coded membrane switches and the
displays are large easy to read liquid crystal type displays (LCD). The
LCD’s are supplied with a backlight.
All units include automatic zero tracking, audible alarm for pre-set
weights, and an accumulation facility that allows the individual weights to
be stored and recalled as an accumulated total.
2. KEY DESCRIPTIONS
Zero or
Set the zero point for all subsequent weighing. The display shows
zero.
A secondary function
other functions.
, of "Enter" key when setting parameters or
Tare or
Tares the scale. Stores the current weight in memory as a tare value,
subtracts the tare value from the weight and shows the results. This is the
net weight. Entering a value using the keypad will store that value as the
tare value. A secondary function ▲, of incrementing the active digit
when setting a value for parameters or other functions.
Units or
Selects the data to be displayed when parts counting and is used for
other functions during setting modes. Will move the active digit to the
when setting values for some functions.
right
The next function is select mode of weight for display, press the key, you
can select gross weight, net weight, tare weight.
-8-
CTB weighing indicator operation manual
-1-
-2F4 off
Smpl or
Select sample quantities when parts counting.
CTB weighing indicator operation manual
REV C1, June 2012
Move the active
digit left
when setting values for other functions.
Func or SET
Used to select the function of the scale. If the scale is weighing it
will select parts counting. Of it is not in weighing mode it will return the
user to weighing. Secondary function ( C ) , is to act as a clear key
when setting values for parameters or other functions.
Print/M+ or
To print the results to a PC or printer using the optional RS-232
interface. It also adds the value to the accumulation memory if the
accumulation function is not automatic. Secondary function (ESC) , is to
return to normal operation when the scale is in a parameter setting mode.
U or Unit
Change weighing unit.
3. BASIC OPERATION
3.1 Zeroing The Display
You can press the ZERO/ENTER key at any time to set the zero
point from which all other weighing and counting is measured, within 4%
of power up zero. This will usually only be necessary when the platform is
empty. When the zero point is obtained the display will show the indicator
for zero.
The scale has an automatic rezeroing function to account for minor
drifting or accumulation of material on the platform. However you may
need to press the ZERO/ENTER key to rezero the scale if small amounts
of weight are shown when the platform is empty.
3.2 Taring
Zero the scale by pressing the ZERO/ENTER key if necessary. The
zero indicator will be on.
CLoCK
bL
bEEP
F5 Prt
Prog
Pin
Set clock off or on,
CLK off
CLK of / CLK on
Set the backlight to be on, automatic or off,
EL Au
EL on / EL Au / EL off
Set the beeper to be
Bp 3
Bp 1, Bp 2 ,BP3 during the check-weighing
function. BP 1: not beeper. BP 2 :If the
weighing
overstep
the
rang
of
check-weighing, it will tweet. BP3 :If the
weight accord the rang of check-weighing, it
will tweet.
Set the RS-232
P Prt: print when the PRINT key is pressed,
P Cont: print continuously, connect to PC , the indicator will
show baud rate (600/1200/2400/4800/9600), then there are
three mode of the cont,
cont 1: it only send gross weight at the weighing mode. at
the counting mode, it will send GW, unit weight and PCS.
Cont 2: it is TOLEDO standard protocol Cont 3: NA
Control command:
Press PC keyboard “T”, the indicator will do tare operate,
the indicator will show net weight. Press PC keyboard “Z” to
rezero the scale if small amounts of weight are shown when
the platform is empty.
Sei re: connect with remote display
ASK: ask and answer mode, PC send data “R” to T2000A,
then T2000A send weighing data back to PC. The function
of “T” and ”Z” is as “P-CONT”.
P auto: auto print
After you setting, display will show “b xxxx”, use TARE key
to select baud rate (600bps~9600bps)
Then display shows “PR x”, set the date/time print format,
Then display show “LAB x”, set gross/acc print format, you
can see detail print out format table below.
Then display show the printer type
TY-TP: mini ticket printer
TY 711: A711 label printer
LP-50: LP-50 label printer
BT-IN: use T2000AP built-in mini printer. (only T2000AP)
Enter the programming and calibration menus by entering
the correct password. See the calibration section for details.
When the scales are set to display in other units of weight the
accumulation function is still keeping the weight in kilograms.
CTB weighing indicator operation manual
REV C1, June 2012
-7Pressing ZERO/ENTER will allow you to set the function. It may be
necessary to either use TARE/+ or set a value using the SMPL/ key
and UNITS/ key to move the active digit and then using the TARE/
key to increment a digit, followed by the ZERO/ENTER key to enter the
value. Use the PRINT/ESC key to leave a parameter unchanged.
For example when the display shows “F0
H-L” press the
ZERO/ENTER key to begin. The display will show “Set Lo”, press the
ZERO/ENTER key to set the low limit, or press the TARE/+ to skip to
the next parameter, “Set Hi” for setting the high limit.
After pressing the ZERO/ENTER key to set a limit, use the SMPL/
key and UNITS/ keys to change the flashing digit, then use the
TARE/ key to increment the flashing digit. Continue to the next digit
and set it as needed.
When all digits have been set press the ZERO/ENTER key to store
the value. The display will go back to the parameter just set, i.e. “Set
Lo”. Advance to another parameter if needed or press the PRINT/ESC
key to return to weighing.
FUNCTION MENU SETTINGS
FUNCTION
F0 H-L
F1 toL
SUBFUNCTION
SEt Lo
SEt HI
to CLr
to P-C
to Prt
F2 unt
F3 tI
SEt dA
SEt tI
DESCRIPTION
Set a value for the Low limit.
Set a value for the High Limit.
Clears the accumulation memory without
printing the results.
Prints the Accumulation memory total and
then clears the memory.
Prints the Accumulation Total, does not clear
the memory.
Sets the displayed second unit of weight.
Select g, Lb, oz , hj or tj.
Set date, The display will show last date set or
00.01.01. Enter new date, format yy. mm. dd
Set time, The display will show current time
Enter new time, format hh. Mm .ss
DEFAULT
VALUE
000.000
000.000
kilogram,
kg
CTB weighing indicator operation manual
-6Place a container on the platform, a value for its weight will be
displayed. Press the TARE key to tare the scale. The weight that was
displayed is stored as the tare value and that value is subtracted from the
display, leaving zero on the display. The "TARE" indicator will be on. As
product is added only the weight of the product will be shown. The scale
could be tared a second time if another type of product was to be added
to the first one. Again only the weight that is added after taring will be
displayed.
When the container is removed a negative value will be shown. If the
scale was tared just before removing the container this value is the gross
weight of the container plus all product that was removed. The zero
indicator will also be on because the platform is back to the same
condition it was when the ZERO/ENTER key was last pressed.
3.3 Weighing a sample. To determine the weight of a sample first tare
the empty container then place the sample in the container. the display
will show the weight and the units of weight currently in use.
4. Parts Counting
When the scale is showing weight, pressing the SMPL key will start
the parts counting function.
Before beginning, tare the weight of any container that will be used,
leaving the empty container on the scale. Place the number of samples
on the scale. The number should match the options for parts counting,
10, 20, 50, 100 or 200 pieces.
Press the SMPL key to begin. The scale will show "SP 10"
asking for a sample size of 10 parts. Change the sample size by
pressing the TARE/ key. the display will cycle through the options:
10,20, 50, 100, 200 and back to 10.
Press the ZERO/ENTER key when the number matches the number
of parts used for the sample.
As more weight is added the display will
CTB weighing indicator operation manual
REV C1, June 2012
-3show the number of parts (pcs).
Press the FUNC key to return to normal weighing.
5. Check-Weighing
Check-weighing is a procedure to cause an alarm to sound when the
weight on the scale meets or exceeds values stored in memory. The
memory holds values for a high limit and a low limit. Either limit can be
used or both can be used.
See PARAMETERS SECTION for the procedure to be used to set
the limits, function "F0 H-L" is used. After limits have been set the
Check-weighing function is enabled.
When a weight is placed on the scale the arrows will show if the
weight is above or below the limits and the beeper will sound as
described below.
BOTH LIMITS SET
The display will show OK and the beeper will sound when the
weight is between the limits.
LOW LIMIT SET,
HIGH LIMIT is set to zero
The display will show OK and the beeper will sound when the
weight is less than the Low Limit. Above the Low Limit the
display will show HIGH and the beeper will be off.
HIGH LIMIT SET,
LOW LIMIT is set to zero
The display will show LOW and the beeper will be off when the
weight is less than the High Limit. Above the High Limit the
display will show OK and the beeper will be on.
BOTH LIMITS SET. LOW IS SET GREATER THAN HIGH
The beeper will never sound and the display will show LOW if the
weight is less that the LOW limit, and HIGH if
CTB weighing indicator operation manual
-4the weight is greater than the Low Limit.
NOTE: The weight must be greater than 20 scale divisions for the
checkweighing to operate.
To disable the Check-Weighing function enter zero into both limits by
pressing the FUNC/C key when the current limits are shown then
pressing ZERO/ENTER to store the zero values.
6. Accumulated Total
The scale can be set to accumulate manually by pressing the PRINT
key. See the PARAMETERS Section for details of selecting the method
using function "F5 PRT". The accumulation function is only available
when weighing. It is disabled during parts counting.
The weight displayed will be stored in memory when the PRINT key
is pressed and the weight is stable.
The display will show "ACC 1" and then the total in memory for 2
seconds before returning to normal. If the optional RS-232 interface is
installed the weight will be output to a printer or PC.
Remove the weight, allowing the scale to return to zero and put a
second weight on. Press the PRINT key, the display will show "ACC
2" and then the new total. Continue until all weights have been added.
To view the totals in memory press PRINT key in zero point.
To clear accumulation memory press enter the PARAMETER
SECTION and use function "F1 TOL". See below.
7. PARAMETERS
The scale has 6 parameters that can be set by the user plus a
method of entering the calibration section.
To set parameters press the FUNC key.
The display will show the first function, "F0 H-L".
Pressing the TARE/+ will cycle through the other functions.
CTB weighing indicator operation manual
REV C1, June 2012
-5-
CTB weighing indicator operation manual