Chapter 6—Preventive Maintenance
Preventive Maintenance—Pump Module
Pump Module
Test Equipment
Test Equipment
Manufacturer
gauge, pressure, digital
(peak hold)
Either of the following:
pressure test
• PTE-1 (accuracy from 0.1 to
• Heise
0.025% span)
(www.heise.com)
• 2089, 2086, or 2084 (accuracy
• Ashcroft
from ±0.05%, 0.10%, or
(www.ashcroft.com)
0.25% of span)
• or an equivalent gauge with:
(a) unit of measure in mmHg
(b) accuracy of ±1%
(c) range of 0-1500 mmHg
hemostat
N/A
N/A
fluid-side occlusion test
IV pole, standard
CareFusion
903-0336 or equivalent
rate, pressure, and occlusion tests
• CareFusion
• 2210-0500
• CareFusion
• CareFusion
• 8100-RCS
• 8100-PCS
IV solution container
(bag preferred)
N/A
N/A
luer lock, female,
1/8"-27 NPT
Cole-Parmer
K-45503-78 or equivalent
(www.coleparmer.com)
pressure test
reducer, female,
1/4" NPT (F) x
1/8" NPT (F)
Cole-Parmer
K-06349-91 or equivalent
(www.coleparmer.com)
pressure test
scale, digital
Acculab
(www.acculab.com)
VIC-212 (VICON Series) or
equivalent with accuracy of
±0.01g
rate and pressure tests
tubing, silicone
CareFusion
303109 or equivalent
pressure test
T-fitting
CareFusion
303815 or equivalent
pressure test
valve, 3-way
CareFusion
97555 or equivalent
pressure test
IV sets
• IV set, standard,
without check valve
• IV set, calibration
• IV set, calibration
Model/Part Number
Application
rate and pressure tests
Alaris® System Maintenance v9.5.X Software User Manual
rate and pressure tests
47