Download Service Manual

Transcript
Performance Tests
Procedure and Equipment
Table 6-1
Required Test Equipment by Model
Agilent Model Number
Model Name
Test Number
Mini-Circuits ZFL-2000
or equivalenta
Amplifier 1
5
GTC RF Products GRF
5016 or equivalentb
Amplifier 2
27, 28
3458A
Multimeter
8-9, 12, 15, 18
5316A
Counter
11, 16
8562A
Spectrum Analyzer
6-7
8663A
Signal Generator (High Performance)
4, 20-23, 28
8648B Option 1EA
Signal Generator
19
8902A
Measuring Receiver
1-5, 17, 19-22, 25, 27
8903B
Audio Analyzer
4, 10, 12, 16, 18, 20-22
11667A
Power Splitter
19, 27, 28
11722A
Sensor Module
5, 19, 25, 27
11793A
Microwave Converter
1-5
E4420B
Signal Generator
1-5, 24, 27
E6380-61811c
System Power Calibration Program Software Kit
19
89441A with options
AYA, AY9, UFG
Vector Signal Analyzer
26
EPM-438
Power Meter
25, 28
8482A
Power Sensor
25, 28
a. Required amplifier specifications are frequency range 1.7 to 2.0 GHz, gain >18 dB, noise
figure <5 dB. For more information about Mini-Circuits, contact them at (718) 934-4500 or
http://www.minicircuits.com.
b. Required amplifier specifications are frequency range 1.0 to 2.0 GHz, gain of 43 dB, output power of +20 dBm. For more information about GTC, contact them at (310) 673-8422
or [email protected].
c. To order the System Power Calibration Card see “Ordering Replacement Parts and Support” on page 85.
143