4 Repair Procedures
4.12.12 Replace
common gas outlet (CGO)
assembly
1. Follow the procedure in Section 4.12.10 to gain access to the CGO.
2. Remove the tubing that is connected to the CGO assembly.
3. Remove the two screws that hold the CGO assembly (A) in the bulkhead.
4. Pull out the CGO assembly.
A
5. Reassemble by following the previous steps in reverse order.
6. Perform the checkout procedure (Section 3).
4.12.13 Remove the
bulkhead
1. Follow the procedure in Section 4.12.10 to gain access to the bulkhead.
2. Remove the canisters and the top canister dish (Section 4.12.7).
3. From underneath, remove the two screws that secure the bulkhead
assembly. (Screws are located near the holes for the upper canister dish
release button.)
4. To reassemble, perform the previous steps in reverse order.
5. Perform the Breathing System Leak Tests (Section 7.2).