Aestiva Anesthesia Machine
Install the molded APL Valve
Actuator
1. Set the APL Valve Actuator assembly to the MIN position.
2. From the top of the control panel, place the molded APL Valve Actuator
into the control panel opening. Align the pointer on the base of the APL
Valve Actuator so that it faces about 30 degrees toward the operator from
a line parallel to the side of the breathing system (or to a customer
preferred position).
3. While holding the APL Valve Actuator in place, swing the control panel to
the fully open position.
4. Place the compression spring over the body of the actuator. For an older
panel (not modified), also place the thin spacer (spiral ring) in the recess
between the spring and body of the actuator.
Extraction Tabs (3)
Ramp
Ramp
Hold-down
Base
Spring
5. Install the hold-down base:
• Position the hold-down base over the actuator so that the extraction
tabs are positioned between the ramps on the end of the actuator.
• Place the heal of your palm over the hold-down base and apply a
slight downward pressure.
• Move your hand backward slightly so that the hold-down base tilts
toward you.
• While maintaining a downward pressure, push your hand forward to
fully snap the hold-down base in place.
4-56
09/06 1006-0452-000