FACSCalibur System User’s Guide 19 Press the air pressure button to turn if off. 20 Press the power button to turn off the Cell Concentrator Module. 21 Drain and fill (PRIME) the flow cell two to three times to remove any air that may have entered the flow cell. 22 Leave the instrument in STNDBY mode with a tube on the SIP containing no more than 1 mL of distilled water. Cleaning the Concentrator Vessel The concentrator vessel should be cleaned and decontaminated after use and before being stored at the end of the day using either 70% ethanol or 10% bleach solution as cleaning solutions. Undiluted bleach contains 5% sodium hypochlorite. m CAUTION: The concentrator vessel and insert holders should not be autoclaved, boiled, or exposed to high temperatures. 127