Download ZVM LIMIT COMPARATOR Operating Instructions

Transcript
LOW LIMIT
LOW LIMIT FAIL LOW LOW LIMIT PASS
CALIBRATION RESISTANCE
RESISTANCE
RESISTANCE [1M - (1M x 10%)]
[1M + (1M x 10%)]
1 Megohm
900Kilohms
1.1 Megohms
HIGH LIMIT PASS
HIGH LIMIT
RESISTANCE
CALIBRATION
[15M? - (15Megohms
RESISTANCE
x 10%)]
HIGH LIMIT
FAIL HIGH
RESISTANCE
[15Megohms +
(15M x 10%)]
15 Megohms
16.5 Megohms
13.5 Megohms
C.) The table below specifies which numbered
resistances on the circuit board correspond to the
optional low limit and the optional high limit. Solder
the appropriate resistors in these places.
D.) Assemble the unit and mark the appropriate places
on the label for the optional limits.
E.) It is recommended to turn the rotary switch to the
optional positions and measure the resistance from
the tip to the body of the plug with an ohmmeter to
verify the resistances.
NOTE: Two (three for position 10) resistor locations are
connected in series are provided in case resistors need
to be added to achieve the desired total resistance.
Solder a shorting wire across the resistor locations that
are not used.
MAINTENANCE
Wipe the plug periodically with alcohol.
Limited Warranty
Vermason expressly warrants that for a period of one (1) year
from the date of purchase, Vermason Limit Comparators will
be free of defects in material (parts) and workmanship (labor).
Within the warranty period, a credit for purchase of replacement
Vermason products, or, at Vermason’s option, the product
will be repaired or replaced free of charge. If product credit is
issued, the amount will be calculated by multiplying the unused
portion of the expected five year life times the original unit
purchase price. Call our Customer Service Department at 0044
(0) 1462 672005 for a Return Material Authorization (RMA) and
proper shipping instructions and address. Please include a copy
of your original packing slip, invoice, or other proof of date of
purchase. Any unit under warranty should be shipped prepaid
to the Vermason factory. Warranty replacements will take
approximately two weeks.
If your unit is out of warranty, Vermason will quote repair
charges necessary to bring your unit up to Vermason factory
standards. Call Customer Service at 0044 (0) 1462 672005 for
proper shipping instructions and address. Ship your unit freight
prepaid.
Figure 2.
Optional low limit
Designated resistors on circuit board
Low limit fail low
R1, R1A (*connected in series)
Low limit pass
R2, R2A (*connected in series)
Optional high limit
Designated resistors on circuit board
High limit pass
R9, R9A (*connected in series)
High limit fail high
R10, R10A, R10B (*connected in
series)
Warranty Exclusions
THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU
OF ALL OTHER PRODUCT WARRANTIES, EXPRESSED AND
IMPLIED, INCLUDING MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY
DISCLAIMED. The express warranty will not apply to defects or
damage due to accidents, neglect, misuse, alterations, operator
error, or failure to properly maintain, clean or repair products.
Limit of Liability
In no event will Vermason or any seller be responsible or liable
for any injury, loss or damage, direct or consequential, arising
out of the use of or the inability to use the product. Before using,
users shall determine the suitability of the product for their
intended use, and users assume all risk and liability whatsoever
in connection therewith.
UNIT C, 4TH DIMENSION, FOURTH AVENUE, LETCHWORTH, HERTS, SG6 2TD UK
Phone: 0044 (0) 1462 672005, • E-mail: [email protected], Internet: Vermason.co.uk
TB-7564 Page 2 of 2
© 2014 Vermason