Contents 4. Troubleshooting 4.1 Fault-Finding Table 11 Common problems and their solution Symptom No output Corrupted output One way communication Unexpected Temperature /Dewpoint readings Unexpected Wind readings Unexpected Temperature/Dewpoint and Humidity readings Unexpected Barometer Readings Status code not 00 Unexpected Rainfall readings or no readings when raining. Solution Check DC power to MetPak RG, cable and connections. Check communications settings of the MetPak RG and host system match, including correct Com port. Check that the unit is in Continuous mode. Check that in-line communication devices are wired correctly. NOTE: It is usual for Anemometer TX + to be connected to converter device RX +. If appropriate use Safe Mode to attempt to obtain communication with the MetPak RG. Check that the communication settings of the MetPak RG and host system match. Try a slower baud rate. Check cable lengths and type of cable. Check that the wiring is in accordance with the manual. Check that the Temperature and Dewpoint units of measure (C, F, K) settings are correct on power up. Check that the Wind Sensor units of measure (m/s, knots, kph, ft/min, mph) settings are correct on power up. Check that the HygroClip filter is clean. Check that the Hub Box Gortex filter is not blocked. See Table 12 Check that the funnel and filter are not clogged and are clean. Check that the tipping bucket mechanism swings freely. MetPak RG User Manual © Gill Instruments Ltd Page 48 Last Change 12 March, 2013