CONNECTING THE TRANSDUCER
The transducer should be connected to the PML 1000 as illustrated in the diagram below. The PML 1000 is
designed to operate with an LVDT in the Secondary to Primary Ratiometric Mode. Using this mode, the PML
1000 determines the measurement by taking into account both the primary and secondary voltages. The PML
1000 measures the differential output of the LVDT (SECA - SECB). The excitation voltage across the primary
coil is also measured. When using this measurement mode the measurements will automatically be
compensated for changes in the excitation voltage. See table on page five for additional LVDT connection
information.
MENU NAVIGATION