Scaling the Display
In order to display the displacement or other parameter in the correct units of measurement it is necessary to
first scale the PML 1000. To do this, select SCLE from the main configuration menu and enter the desired Low
and High display values. The low display value corresponds to the displacement at the Low input (calibration)
point i.e. 0 (mm). The high display value corresponds to the displacement at the high input (calibration) point i.e.
200 (mm).
Calibration
Subsequent routine calibrations of the LVDT to the PML 1000 can be performed either from within the menu's or
by assignment of either a status input or function key. The low and high display scaling values will be retained
throughout subsequent calibrations, all subsequent calibrations must be performed using the same points
otherwise the LDSP and HDSP values will need to be changed.
The calibration points used may be set anywhere in the display range of the PML 1000 i.e. -19999 to 99999.
The can read above and below the nominal calibration points although for best performance it is advisable to
calibrate at the extremes of the required measurement range.
ORDERING INFORMATION
NORTH AMERICA
EUROPE
Measurement Specialties
1000 Lucas Way
Hampton, VA 23666
United States
Tel: 1-800-745-8008
Fax: 1-757-766-4297
Sales: [email protected]
Measurement Specialties
(Europe), Ltd.
Hauert 13
44227 Dortmund
Germany
Tel: +49 (0) 231 9740 0
Fax: +49 231 9740 20
Sales: [email protected]
ASIA
Measurement Specialties
(China), Ltd.
No. 26 Langshan Road
Shenzhen High-Tech Park (North)
Nanshan District, Shenzhen 518107
China
Tel: +86 755 3330 5088
Fax: +86 755 3330 5099
Sales: [email protected]
The information in this sheet has been carefully reviewed and is believed to be accurate; however, no responsibility is assumed for
inaccuracies. Furthermore, this information does not convey to the purchaser of such devices any license under the patent rights to the
manufacturer. Measurement Specialties, Inc. reserves the right to make changes without further notice to any product herein. Measurement
Specialties, Inc. makes no warranty, representation or guarantee regarding the suitability of its product for any particular purpose, nor does
Measurement Specialties, Inc. assume any liability arising out of the application or use of any product or circuit and specifically disclaims
any and all liability, including without limitation consequential or incidental damages. Typical parameters can and do vary in different
applications. All operating parameters must be validated for each customer application by customer’s technical experts. Measurement
Specialties, Inc. does not convey any license under its patent rights nor the rights of others.
APP-01025, Rev A
PML 1000 Operating Instructions
www.meas-spec.com
18/18
1109