Chapter 4, Description of Fields Fields Screens Where Field is Present CALL CONTROL Pwr Up Reg This field determines if, at power-up, the mobile station will attempt to register with the Test Set. • • On - the mobile station will perform a power-up registration Off - the mobile station will not perform a power-up registration GPIB Example “CDMA:CELL:CONF:PUR ‘On’"“ causes the mobile station to attempt to register at power-up Operating Considerations This field provides the option of performing a zone-based registration when a mobile station finds service (Register (CDMA) field description, on page 358) instead of waiting for the power-up registration. This field should be set to the desired mode before a mobile station finds CDMA service. When this field is “On” the Test Set instructs the mobile station to attempt a registration upon power-up. This instruction is sent as part of the paging channel overhead message System Parameters, and is received by the mobile station soon after the mobile station finds service. See Also “Setting up a Call” chapter in the E8285A Application Guide Screens Where Field is Present CDMA CELL SITE CONFIGURATION 349