Download Agilent Technologies N8201A Technical data

Transcript
24
System
Time/Date
Displays the Time/Date function menu keys used to set and display the real-time clock.
Key Path:
System
Remote Command:
There is no remote command for this key.
Time/Date
Turns the display of the real-time clock on or off.
Key Path:
System, Time/Date
Annunciation/Annotation:
of the screen.
State Saved:
Current time and date are displayed in the upper-left corner
Survives Preset and power cycle, but not saved in Instrument State.
Factory Default:
On (Restored by System, Restore Sys Defaults.)
Remote Command:
:DISPlay:ANNotation:CLOCk[:STATe] ON|OFF
:DISPlay:ANNotation:CLOCk[:STATe]?
Example:
DISP:ANN:CLOC ON
Date Format
Enables you to set the date display to month-day-year or day-month-year. It is set to a
month-day-year format when the instrument System Defaults are restored. This key only
affects display of date at the top of the screen, not in the file catalog.
Key Path:
System, Time/Date
State Saved:
Survives Preset and power cycle, but not saved in Instrument State.
Factory Default:
MDY (Restored by System, Restore Sys Defaults.)
Remote Command:
:DISPlay:ANNotation:CLOCk:DATE:FORMat MDY|DMY
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz