Transcript
Section 5: Startup and meter configuration August 2014 Reference, Installation, and Operations Manual 3-9000-743 Rev U 2. Choose a Flow Calibration Method: • Polynomial • Piece-wise linear • None 3. Enter the appropriate forward and reverse meter factors and click the Write button to send the factors to the meter. 4. Click the Cancel button to close the dialog box. If you made changes, click the Write button to apply the changes. After the meter is configured, set S-2 Position 4 (on the CPU Board) back to the open position to write-protect the configuration. 164 Set calibration parameters