8712ET/ES and 8714ET/ES
RF Network Analyzers
Performance Checks
System Performance
Table 1-4 System Performance, Corrected, 50Ω, T/R Calibration
8712ES/8714ES (50 Ω)
85032B/E (Type-N, 50 Ω) Cal Kit, T/R Calibration
Specification1 (in dB)
Description
Directivity
300 kHz
to 1.3 GHz
1.3 GHz
to 3 GHz
50
47
Reflection (One-Port Cal)
42
36
Transmission (Enhanced Response Cal)
42
36
Transmission (Response Cal)
18
15
Load Match
18
15
Reflection Tracking
±0.02
±0.02
Enhanced Response Cal
±0.040
±0.055
Response Cal
±0.17
±0.3
Source Match:
Transmission Tracking:
1. These specifications apply for measurements made using
“fine” bandwidth, no averaging, and at an environmental
temperature of 25° ± 5 °C, with less than 1 °C deviation
from the calibration temperature.
Service Guide