Download Agilent Technologies Series N6700 Specifications

Transcript
Calibration
3
Calibration Set-up
The following figures show the calibration test set-ups. Connect all
leads to the output terminals as shown in the following figure. Twist
the + and - load leads to reduce noise and inductance.
+S
+
-S
+S
+
-S
+S
+
+S
-S
+
-S
+
Agilent 3458A
configured as a
voltmeter
+
Agilent 3458A
configured as a
voltmeter
A.
Current
shunt
25 Ohm resistor
(not used for Models
N6753A & N6754A)
+
Agilent 3458A
configured as
an ammeter
+S +LS -LS -S
B.
D.
C.
+S
+
-S
+
Agilent 3458A
configured as a
voltmeter
Aux Meas+
-
-
G
+ +
+s G -s
Aux Meas-
-
-
G
+ +
sense jumpers
+s G -s
+
+
DC
voltmeter
sense jumpers
+
+
Electronic load
E.
F.
NOTE
Series N6700 Service Guide
Current
monitor
External
DC supply
G.
Agilent 3458A
configured as a
voltmeter
Load lead lengths for models N678xA SMU must be between 1 and 4 feet.
45