Download Mantis 5000 Specifications
Transcript
An integrating sphere is sensitive to small-angle scatter from the sample beam coupling optics. Sometimes, the scattered radiation strikes the wall of the integrating sphere near the reflectance sample port, creating a "halo" surrounding the port. This halo-effect causes a small error in the measurement of the reflectance factor that is most significant when measuring samples of very low reflectance. This error is easily corrected by using the light trap zero baseline correction method discussed above. Simple Baseline Scan for Reflectance or Transmission Measurements Baseline scan data is recorded from the Cary WinUV Scan application. Prior to recording the scan, make sure that mirror M3 is located in the correct position. To collect a baseline scan: 1. Click the Setup button and set up instrument parameters as required by the intended application. Select the %R or %T ordinate scale, depending on the type measurements anticipated. The Beam mode should be double. 2. Select the Baseline radio button on the Baseline tab. 3. Load a Spectralon standard at the reference port and a calibrated reflectance standard, if available, at the sample reflectance port. If a calibrated standard is unavailable for reflectance measurements, an uncalibrated Spectralon standard can be substituted. Install the magnetic port covers over their respective ports. 4. Click the Baseline button. The recorded baseline correction will be effective until replaced by a follow-up baseline scan. Zero Baseline Scan for Reflectance or Transmission Measurements Prior to recording the scan, make sure that mirror M3 is located in the correct position. To record a zero baseline scan: 1. Click the Setup button and set up instrument parameters as required by the application. Select the %R or %T ordinate scale, depending on the type measurements anticipated. The Beam mode should be double. 2. For reflectance measurements, select the Zero/baseline correction or Zero x std ref correction radio button on the Baseline tab. For transmission measurements, Zero/ baseline correction should be selected. 3. Load a Spectralon standard at the reference port and a calibrated reflectance standard, if available, at the sample reflectance port. If a calibrated standard is unavailable for reflectance measurements, an uncalibrated Spectralon reflectance standard can be substituted. Install the magnetic port covers over their respective ports.