Download Mantis 5000 Specifications
Transcript
Introduction The traditional use of the spectrophotometer is to measure the absorbance or transmission of a clear or translucent liquid or solid. Typically UV/VIS/NIR spectroscopy applications include reaction kinetics, quantitative analysis or the identification of the chemical constituents in a substance. The measurement of reflectance offers an added dimension to the spectroscopy capabilities of your spectrophotometer, the analysis is no longer limited to the portion of the beam that penetrates the sample. Accordingly, the analysis is no longer limited to samples that are clear or translucent. The DRA provides the means of measuring absorbance or transmission of opaque, turbid and reflecting substances. Figure 1. The external DRA. Figure 2. The external DRA viewed from the back. There are three versions of the diffuse reflectance accessory that operate over different wavelength ranges.These are specified below. While all of the following variations are available, the configurations indicated by ticks in the table below offer optimal operation. (External DRA 900) (External DRA 1800) (External DRA 2500) Cary 4000 Cary 5000 Cary 6000i The Varian 900, 1800, 2500 External Diffuse Reflectance accessory (DRA) consists of a 150 mm diameter integrating sphere. The sphere is easily installed in the sample compartment of the instrument and features an inbuilt high performance photomultiplier tube (PMT). The DRA also has a Lead Sulfide (PbS) or InGaAs detectors which become active Varian 2500 External DRA. The coating is Polytetrafluoroethylene (PTFE), which