Download ZeissS S81 Technical data

Transcript
40
New installation of S8 / S 81 ceiling mount
Checking the power connection
Caution:
The ceiling mount can be set at two points to the rated voltage used in the
country of destination:
e
Transformer terminal on interface plate.
Settings: 100 - 115 - 230 V
e
Power supply unit
Settings: 115 - 230 V
The correctness of the factory-adjusted setting is checked by our service
team during first-time installation.
e
The power connection must conform to the regulations applicable in
the country of destination.
e
A power switch (2-pin) has been provided.
e
Line fuse
230 V: min. 10 A, slo-blo, class C
115 V: min. 20 A, slo-blo, class C
e
All cables and plugs must be in perfect condition.
e
Potential equalization: if required, the instrument can be incorporated
into protective "potential equalization" measures.
e
A connector for OR illuminators is provided (option).
We recommend integrating the power switch in the central OR control
panel.
M-30-1382-en
S8 / S81 Ceiling Mounts
Issue 12.0