Download ZeissS S81 Technical data
Transcript
71 Accessory interface Electrical connections for accessory modules The power supply for the electrical accessory components can be connected to connectors "A" on the interface plate (connectors for OR illuminators, terminal blocks B and C). On-site requirements for the connection of accessories g Route the power lines in the building up to the interface plate. g Route any required control lines and/or monitor connections in the building up to the interface plate. g Make sure (for in-wall installation) that a sufficient number of conduits with sufficient cross section are provided. g Route additional lines for potential equalization to the terminals on the interface plate. Warning! e The planning, execution and inspection of the on-site electrical installations must be performed by expert electrical technicians and officially registered electrical firms. M-30-1382-en e The electrical installations of the room concerned must comply with the requirements of the applicable national regulations. In the Federal Republic of Germany, VDE 0107 applies. e The number of circuits is dependent on the customer's specific carrier system configuration. e If several circuits are required by the customer, appropriate allowance must be made on the interface plate. e Multiwire lines must be provided with wire end sleeves. S8 / S81 Ceiling Mounts Issue 12.0 Printed on 19. 11. 2007