Download encoder position display unit
Transcript
Measured Value Output Measured value output when probing Parameter P96 allows you to activate measured value output when probing with the KT edge finder. The edge finder is connected to D-sub input X10. Whenever the Probe Edge function is used, your display unit outputs the position of the edge in the selected axis and the actual positions of the other axes over the TXD line of the RS-232-C/V.24 interface. Whenever the Probe Midpoint function is used, your display unit outputs the calculated midpoint in the selected axis and the actual positions of the other axes. Measured value output with CTRL B is inhibited when a probing function is active. Delay times with data output Duration of the latch signal: Storage delay: Data output after: Regeneration time: te ≥ 4 µs t1 ≤ 4.5 ms t2 ≤ 50 ms t3 ≥ 0 Duration of data output in seconds: tD = 176 x number of axes + 11 x number of blank lines Baud rate Next possible signal for measured value output: tE = t1 + t2 + tD + t3 [s] 56 te te t1 t2 t3 tD