Download 1575 1590 Users Guid..
Transcript
11 2590 Scanner (optional) Scanner Operation of sequence, the number of channels in the sequence, and the number of measurements to make on each channel in the sequence. The mode determines the type of sequence to use. You can select among the following modes. Mode Channels SCAN scanner 1 - scanner 2 - scanner 3 ... scanner 10 IN2+SCAN input 2 - scanner 1 - scanner 2 ... scanner 10 ALT IN2/SCAN input 2 - scanner 1 - input 2 - scanner 2 ... scanner 10 ALT SC1/SCAN scanner 1 - scanner 2 - scanner 1 - scanner 3 ... scanner 10 USER-DEFINE arbitrary sequence For the four pre-defined sequences, the CHANNELS parameter sets the number of scanner channels, or in other words, the last scanner channel number in the sequence. For the user defined sequence, CHANNELS sets the total number of steps in the sequence. After the thermometer measures the last channel, the sequence repeats starting again at the beginning. The NUMBER OF SAMPLES parameter specifies the number of measurements to make on each channel before advancing to the next. The scanner sequence mode works with either of the sampling modes described in Section 7.2. If continuous sampling is selected, the channels will sequence indefinitely. If SAMPLE N is selected, the channels will be sequenced until the total number of measurements reaches the number specified with SAMPLE N. 131