Download MW100 Viewer Software User`s Manual
Transcript
4.2 Calibration Procedure Notes on Calibration • Do not perform other operations during calibration. Take extra care because if you perform another operation while calibration is in progress, the module may suffer a break down. • When the screw terminal block (model 772080) is connected to the 10-CH MediumSpeed Universal Input Module, the terminal arrangement differs from that of clamp terminals. With the calibration software, modules with screw terminal blocks connected are recognized as clamp terminals, meaning that the wiring diagrams show the wiring for the clamp terminals. Therefore when wiring and performing calibration, you must follow the markings showing the terminal functions and the terminal codes indicating the types of signals input to each terminal located on the back of the terminal cover of the 10-CH screw terminal block. Screw terminal block (model: 772080) terminal wiring markings. 1 CH6 2 1 CH1 2 3 CH7 4 3 CH2 4 5 CH8 6 5 CH3 6 7 CH9 8 7 CH4 8 9 CH10 10 9 CH5 10 11 RTD 11 RTD Note: The b terminal is common to all channels. The b terminals are connected internally. Procedure 1. Click the illustration of the input/output module that you wish to calibrate. 4-4 Moving the cursor over an module that can be calibrated shows the word “Calibrate” IM MW180-01E