WARNING Do: Store sensor between 36° F-77° F during its shelf life. Why: Storing the sensor incorrectly might cause the sensor glucose readings to be inaccurate. Never store sensors in the freezer. Consequences: If stored outside of 36° F-77° F, your sensor glucose readings may not be accurate, resulting in you missing a severe low or high glucose event. WARNING Don’t: Never use the Dexcom G5 Mobile CGM System sensor glucose readings for treatment decisions. Do: Only use BG values from your BG meter for treatment decisions (e.g., how much insulin to take). Why: Since they measure your glucose from different body fluids, sensor glucose readings can be different from your meter’s blood glucose values. Consequences: Using sensor glucose readings for treatment decisions could result in you missing a severe low or high event. PRECAUTION Do: Select sensor insertion site with care. Avoid: • Areas likely to be bumped, pushed or squeezed • Areas of skin with scarring, tattoos, or irritation • Injecting insulin within 3 inches of sensor • Placing an insulin pump infusion set within 3 inches of sensor Why: Inserting sensor in these areas may affect sensor glucose readings. Consequences: Inaccurate sensor glucose readings may result in you missing a severe low or high event. Dexcom G5 Mobile System User Guide Starting a Sensor Session: Inserting Sensor, Transmitter, and Starting Your Session 71